Skip to content

Public FDA record

Z-0633-04: Bravo pH Monitoring System Receiver

Recall record Z-0633-04 for Medtronic Gastroenterology / Urology. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0633-04
Company in FDA record
Medtronic Gastroenterology / Urology
FDA classification date
Jul 20, 2004
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

The product's rear panel label could smear during use causing the serial number to be unreadible. This has the remote possibility of resulting in patient data from diagnostic tests being mixed-up or attributed to the wrong patient.

Source facts

Action
Letters dated February 6, 2004 were sent to the consignees. Enclosed with the letter to each consignee were label sticker(s) having the serial numbers of the device(s) purchased by the consignee. The letters instruct consignees to place the enclosed stickers on the devices over the previous back panel labels if the serial number of the label matches the serial number on the device. Special instructions for devices with smeared labels include the return of devices with unreadable serial numbers.
Initiated
2004-02-06
Root cause
Other
Terminated
2005-11-21
Distribution
The devices were distributed nationwide in the United States and worldwide.
Recall number
Z-0633-04
Recall status
Terminated
CDRH record ID
31446
FDA classified
2004-07-20
Recall event number
28253

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗