Open original FDA record ↗Recall record
- FDA identifier
- Z-0634-2021
- Company in FDA record
- Roche Diagnostics Operations, Inc.
- FDA record creation date
- Dec 17, 2020
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Roche confirmed customer complaints of low quality (QC) recovery and invalid calibrations for the Tina-quant C-Reactive Protein Gen. 3 on the cobas c701 and 702 modules. A discoloration (yellow color) was observed and is the result of cross contamination between the R1 and R3 reagents.
Source facts
- Action
- On November 6, 2020, the firm mailed customers Urgent Medical Device Correction letters. Customers were informed that the present communication serves as an update to previous UMDCs sent on 9/11/2019, 12/18/2019, and 8/27/2020, in that it includes the new product and lot number listed above. Customer Instructions: " Discontinue use and discard any remaining affected product in your inventory according to your local waste guidelines. " Refer to the Product Replacement section on page 2 of the UMDC for product replacement information. " Complete all sections of the enclosed faxback form (TP-01128) and fax or email it according to the instructions on the form even if you are not requesting product replacement. " If your facility has distributed the affected product to another site, please ensure this UMDC is provided to that site. " File this UMDC for future reference. Please contact the Roche Support Network Customer Support Center 24 hours a day, seven days a week at 1-800-428-2336 if you have questions about this recall.
- Initiated
- 2020-11-06
- Root cause
- Process control
- Terminated
- 2024-10-18
- Distribution
- US Nationwide distribution including in the states of AL, CA, FL, GA, HI, IA, IL, IN, KY, MA, MD, MI, MN, MO, NJ, NY, OH, PA, SC, TN, TX, VA.
- Recall number
- Z-0634-2021
- Recall status
- Terminated
- CDRH record ID
- 184441
- FDA record created
- 2020-12-17
- Recall event number
- 86745