Open original FDA record ↗Recall record
- FDA identifier
- Z-0634-2022
- Company in FDA record
- Synthes (USA) Products LLC
- FDA record creation date
- Feb 10, 2022
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Top cap may loosen and detach if the Universal Chuck becomes jammed and the user attempts to manually free the device, internal ball bearings may fall out onto the surgical field and cause surgical delay or, and if not recovered, adverse tissue reaction and infection.
Source facts
- Action
- Depuy Synthes issued Urgent Medical Device Recall (Removal) issued on 12/22/21 via email FedX. Letter states reason for recall, health risk and action to take: 1. Examine your inventory immediately to determine if you have the subject products and quarantine the subject product. DO NOT USE THE SUBJECT PRODUCT. 2. Contact your DePuy Synthes Sales Consultant or call the customer support services at 1-800-479-6329 to coordinate the return/credits of the subject product. 3. Review, complete, sign, and return the attached Business Response Form (page 3 of this letter) to OneMD-Field-Actions@its.jnj.com within 3 business days of receipt of this notification. Please include in the email subject: FA 2050712: 2021 Universal Chuck. " Please complete the attached Business Response Form even if you do not have the subject product on hand. 4. Forward this notice to anyone in your facility that needs to be informed (i.e., those who manage, transport, store, stock, or use the subject products). 5. If any of the subject product has been forwarded to another facility, contact that facility and provide them with this notice. 6. Post a copy of this notice in a visible area for awareness and keep a copy for your records. Should you have any inquiries please do not hesitate to contact your DePuy Synthes Sales Consultant. For Medical Information request, please visit our website: https://www.jnjmedicaldevices.com/mir
- Initiated
- 2021-12-22
- Root cause
- Device Design
- Distribution
- US Nationwide distribution.
- Recall number
- Z-0634-2022
- Recall status
- Open, Classified
- CDRH record ID
- 191533
- FDA record created
- 2022-02-10
- Recall event number
- 89428