Open original FDA record ↗Recall record
- FDA identifier
- Z-0635-2021
- Company in FDA record
- Helena Laboratories, Corp.
- FDA record creation date
- Dec 17, 2020
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Due to complaints received associated with cracked/split test tube caps.
Source facts
- Action
- On 11/09/2020, international distributors were initially contacted and provided with a notification for distributors letter. On 11/16/2020 domestic customers were contacted via telephone and then a formal notification was delivered. The Recalling Firm was informing its customers to the potential of defect associated with the (flip-top) caps for some test tubes where the cap my split resulting in leaking of biohazardous fluids as users are instructed to shake the tube for proper mixing. Customers are instructed to immediately examine their inventory and quarantine affected product. If customers have further distributed the affected product, they are asked to identify the users within their facility to notify them at once of this product recall. Customers are also asked to complete and return the response form as soon as possible. For any questions, customer can email or call the Recalling Firm as jhanka@Helena.com or 1-800-231-5663 extension 1177.
- Initiated
- 2020-11-09
- Root cause
- Nonconforming Material/Component
- Distribution
- Worldwide distribution - U.S. Nationwide including in the states of NJ, NY, PA and TX. The countries of Austria, Canada, England, France, Italy, Portugal, Russia, Spain.
- Recall number
- Z-0635-2021
- Recall status
- Open, Classified
- CDRH record ID
- 184497
- FDA record created
- 2020-12-17
- Recall event number
- 86774