QC testing of the Antibody Reagent Set for Lp(a) SPQ Test System it was discovered that the background was failing high of the specification. Through substitution testing it was determined that the Lp(a) diluent, part #238017 was the cause of the failing background specification.
Source facts
Action
Customers were sent notification letters and instructed to destroy any remaining inventory. Customers were also requested to return a form which indicates number of kits destroyed and a replacement was offered.
Initiated
2005-01-10
Root cause
Other
Terminated
2006-07-22
Distribution
KC, MO, CA, IN, HI, GA, TX, MN, ME, AZ, FL, VA, MI, NJ, NC, NY, CO, WI, MS, BRUSSELS, GREECE, SWEDEN, ENGLAND, SPAIN, CANADA & SWITZERLAND
Recall number
Z-0636-05
Recall status
Terminated
CDRH record ID
37152
FDA classified
2005-03-18
Recall event number
31104