Z-0636-2013: Siemens CLINITEK Atlas Positive and Negative Control Strips for Urine Chemistry Catalog Number: 5019 Whe

Recall record Z-0636-2013 for Siemens Healthcare Diagnostics Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0636-2013
Company in FDA record
Siemens Healthcare Diagnostics Inc
FDA classification date
Jan 4, 2013
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 17, 2026
Recall classification
Class III

Reason stated in the source

When these lots of control material are reconstituted, the positive control solution leukocyte result may become negative if the solution is held for more than two hours above 25C (77F).

Source facts

Action
Siemens issued An Urgent Field Safety Notice via e-mail to all affected SiemensHealthcare Diagnostics Regional offices on November 21, 2012 both in the United States and foreign accounts. This notice informed customers of the issue with the CLINITEK Atlas and Chek-Stix Positive Urinalysis Control Solutions When these lots of control material are reconstituted, the positive control solution leukocyte result may become negative if the solution is held for more than two hours above 25C (77F). You can use the reconstituted positive control solution for up to eight hours if the solution is held at 25C (77F) or lower. If the operating temperature is greater than 25C and lower than 30C (77F and 86F), the solution can be used for up to two hours. For questions regarding this recall call 781-269-3158.
Initiated
2012-11-21
Root cause
Nonconforming Material/Component
Terminated
2014-11-06
Distribution
Worldwide Distribution - USA (nationwide) and internationally to Canada, Aruba, Brazil, Denmark, India, Mexico, New Zealand, Singapore,Thailand, Bahamas, South Wales, and Hong Kong,
Recall number
Z-0636-2013
Recall status
Terminated
CDRH record ID
114818
FDA classified
2013-01-04
Recall event number
63809

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗