Open original FDA record ↗Recall record
- FDA identifier
- Z-0637-2021
- Company in FDA record
- Boston Scientific Corporation
- FDA record creation date
- Dec 17, 2020
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
This is a follow-up to the EKOS Field Safety Notice dated April 30, 2019. The Radio Frequency (RF) board has been redesigned to prevent occasional internal board failures at lower temperatures. All EKOS CU 4.0 consoles containing the old RF board design will be replaced with the newly designed CU 4.0 consoles.
Source facts
- Action
- On November 3, 2020, the firm distributed Urgent Medical Device Recall letters to affected customers. Customers were asked to do the following: Interim Recommendations until your EKOS CU4.0 console has been exchanged: In order to avoid the channel errors described in this notice: 1- Store the Control Unit at room temperature, in a well-ventilated area. 2- The console should be plugged in and powered on for 30 minutes prior to making connections and starting therapy to allow the Control Unit to warm up to operating temperature (+15C to +40C) Customers were also asked to complete and return the verification form to their local Boston Scientific office. The recalling firm will provide affected accounts with an updated CU 4.0 unit with a newly designed RF board.
- Initiated
- 2020-11-03
- Root cause
- Device Design
- Distribution
- No US distribution. Foreign distribution to Latin America and EU.
- Recall number
- Z-0637-2021
- Recall status
- Open, Classified
- CDRH record ID
- 184677
- FDA record created
- 2020-12-17
- Recall event number
- 86847