An internal inspection indicates that Calibrator 2 and Conjugate are switched in their positions in the integral. This will result in Calibrator 1 RLU's similar to that of background RLU, and Calibrator 2 RLU's near expected range. A curve can not be calculated from this data.
Source facts
Action
DiaSorin Customer Notification dated 02/14/05 was sent to customers instructing them to destroy all kits from the affected lot number. Customers were also asked to complete and return a form indicating how many kits were destroyed.
Initiated
2005-02-14
Root cause
Other
Terminated
2006-07-22
Distribution
MA, Spain, Germany & Belgium
Recall number
Z-0638-05
Recall status
Terminated
CDRH record ID
37417
FDA classified
2005-03-18
Recall event number
31228