Open original FDA record ↗Recall record
- FDA identifier
- Z-0638-2018
- Company in FDA record
- Zimmer Biomet, Inc.
- FDA record creation date
- Feb 16, 2018
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Robot arm being sent to the wrong position
Source facts
- Action
- Urgent Medical Device Recall (Correction) notices were sent to customers. The letter provided to hospital risk managers and surgeons identified the issue and their responsibilities. These responsibilities included: - Stop using the micromovement function in isocentric mode during Brain Surgery with ROSA Brain 3.0 device. -Reviewing the notification to ensure comprehension of the steps required to be taken prior to the permanent correction that will be implemented on site by a service engineer. - Completing and returning Attachment 1 (Certificate of Acknowledgement) to the e-mail address provided. The firm has prepared a software correction that will be deployed and implemented on site by a MEDTECH representative.
- Initiated
- 2017-05-01
- Root cause
- Software design
- Terminated
- 2020-07-13
- Distribution
- Distributed to CA, DC, FL, IL, KY, MA, MN, NC, NY, OH, and PA.
- Recall number
- Z-0638-2018
- Recall status
- Terminated
- CDRH record ID
- 160939
- FDA record created
- 2018-02-16
- Recall event number
- 78944