Z-0638-2022: OER-Elite Endoscope Reprocessor-for use in cleaning and high-level disinfection of heat sensitive Olympus

Recall record Z-0638-2022 for Olympus Corporation of the Americas. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0638-2022
Company in FDA record
Olympus Corporation of the Americas
FDA record creation date
Feb 11, 2022
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

The detergent and alcohol tank tubing may have been incorrectly assembled in the OER-Elite. If the tubing is incorrectly assembled, alcohol will be dispensed instead of detergent for the cleaning cycle and detergent will be dispensed instead of alcohol for the alcohol cycle, resulting in a ineffective endoscope reprocessing and patient exposure to residual detergent

Source facts

Action
Olympus issued Urgent Medical Device Correction letter dated 12/13/21. Letter states reason for recall, health risk and action to take: 1. Please immediately perform an inspection of your installed OER-Elite units as described in the Attachment 1 Alcohol Rinse Inspection . If you require assistance, please contact your local Endoscopy Support Specialist (ESS). 2. If the inspection check does not pass, stop using the OER-Elite and endoscopes reprocessed with that OER-Elite. Contact Olympus TAC at 1-800-848-9024, option 1 for further instruction. If the inspection check passes, you may continue to use your OER-Elite. 3. Access the Olympus recall portal to indicate that you have received this notification and to report the results of the inspection. Go to https://olympusamerica.com/recall. Enter the recall number "0404" and provide your contact information as indicated in the portal. An Olympus Endoscopy Support Specialist (ESS) will reach out to you to schedule an on-site inspection of the OER-Elite detergent and alcohol tubing assembly for final confirmation. If you do not know your local ESS, please contact our Olympus TAC at 1-800-848-9024 to connect you with your local ESS. In addition, if you have further distributed this product, please reach out to those facilities and notify them at once of this field action and appropriately document your notification process. Your notification to your customers may be enhanced by including a copy of this field action notification letter. If you require additional information, please do not hesitate to contact me at Cynthia.Ow@Olympus.com.
Initiated
2021-12-13
Root cause
Process control
Terminated
2023-12-26
Distribution
Nationwide
Recall number
Z-0638-2022
Recall status
Terminated
CDRH record ID
191488
FDA record created
2022-02-11
Recall event number
89404

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗