Spinal fixation system 7.5mmx 50mm screws are packaged with the correct labeling, but the screws themselves were laser marked with an incorrect size description of 7.5 x 40; use of incorrect screws could lead to screw perforation and indention of vascular or other structures.
Source facts
Action
On 10/16/23, recall notices were sent to customers asking them to do the following:
1) Locate, quarantine, and return affected screws to the recalling firm.
2) Complete and return the inventory return certificate form via email to Alex.Hayden@ZimVie.com
Recall-related questions or concerns can be directed to please email Alex.Hayden@ZimVie.com.
Initiated
2023-10-16
Root cause
Nonconforming Material/Component
Distribution
US Nationwide distribution in the states of MN, NE.
Recall number
Z-0639-2024
Recall status
Open, Classified
CDRH record ID
205050
FDA classified
2024-01-04
Recall event number
93650
Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.
Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.