Z-0640-2022: Liko Mobile Lifts: Primarily intended for use in nursing homes LikoM230 Model: 2050015

Recall record Z-0640-2022 for Hill-Rom, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0640-2022
Company in FDA record
Hill-Rom, Inc.
FDA record creation date
Feb 16, 2022
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Missing bushing on Liko M220 and M230 causing wear and potential for patient falling

Source facts

Action
HillRom issued Urgent Medical Device Correction Letter (FSCA Identifier: FA 2021-010-002-LUL-001) via third party on 23 December 2021 via UPS Ground Service to Distributors and Direct Users. Letter states reason for recall, health risk and action to take: 1. Review the product label on your Liko M220/M230 to determine if it is impacted by this field action by identifying the manufacturing date of the device (i.e. Manufacturing date between 2016-DEC-27 and 2021 SEP-01). See figure 2. 2. Please periodically inspect your Liko device for damage or wear. Report visible wearing or damage to your Technical Support Team do not use the device until Technical Support can inspect and repair where necessary. 3. Please fill out the attached response form and return it to Hillrom hillromLUL001@sedgwick.com within two weeks Please share this Field Safety Notice with your end users and complete the attached response form and return to hillromLUL001@sedgwick.com within two weeks. Contact hillromLUL001@sedgwick.com to receive an electronic copy of this notification and response form for onward distribution. Action to be taken by the Hillrom: Upon receipt of the response form, identifying potentially impacted devices, Hillrom or an official Hillrom distributor/representative will be in contact to schedule inspection of your device to confirm presence of the bushing. Where it is confirmed the device is missing the bushing, a replacement slingbar with the bushing will be installed. If you have any questions regarding this safety notice, please contact Hillrom Technical Support at 800-445-3720 or technical.support@hillrom.com
Initiated
2021-12-23
Root cause
Process control
Distribution
Worldwide distribution - US Nationwide and the countries of Austria, Austria, Bahrain, Canada, Switzerland, Germany, Denmark, Spain, Finland, France, United Kingdom, Greece, Hong Kong, Ireland, Israel, Italy, Norway, Saudi Arabia, Sweden, Slovenia, Taiwan (Province of China), South Africa, Oman.
Recall number
Z-0640-2022
Recall status
Open, Classified
CDRH record ID
191505
FDA record created
2022-02-16
Recall event number
89414

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗