- FDA identifier
- Z-0641-05
- Company in FDA record
- Heartsine Technologies Inc
- FDA classification date
- Mar 23, 2005
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
The affected Automatic External Defibrillators (AED) annouce a fault warning and shut themselves off before a shock can be delivered to the patient. This is reportedly due to slow capacitor charging rates which are interpreted by the AED software as a fault condition.
Source facts
- Action
- Recall was initiated by telephone 2/14/2005 with follow-up fax, email or postal mailing using one of several customized letters for distributors or customers. Customers will be sent an upgrade kit by overnite mail and forms to confirm upgrade.
- Initiated
- 2005-02-14
- Root cause
- Nonconforming Material/Component
- Terminated
- 2011-06-08
- Distribution
- Nationwide to distributors and user facilities. Also to Argentina, Austria, Belgium, Finland, France, Germany, Hong Kong, Ireland, Israel, Italy, Japan, Kuwait, Netherlands, Portugal, Puerto Rico, Singapore, Spain, Switzerland, Thailand, United Kingdom.
- Recall number
- Z-0641-05
- Recall status
- Terminated
- CDRH record ID
- 37385
- FDA classified
- 2005-03-23
- Recall event number
- 31212