Open original FDA record ↗Recall record
- FDA identifier
- Z-0641-2022
- Company in FDA record
- Microbiologics Inc
- FDA record creation date
- Feb 14, 2022
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class III
Reason stated in the source
Some of the KWIK-STIK Catalog #01023P Campylobacter coli, Lot# 1023-05-7, contained an inner label stating the product is Catalog #0919K Haemophilus influenza. The organism has been confirmed to be C. coli, indicating this is a labeling error and the organism is the correct organism.
Source facts
- Action
- Microbiologics notified customers on 03/25/2013 via letter titled "Important Medical Device Correction". Instructions included to inspect inventory for affected products. If customers have received the product identified within the recall (KWIK-STIK" 01023P, Lot# 1023-05-7), it is recommended that the customer destroys the product. Microbiologics shall replace the destroyed product upon confirmation of destruction. Questions or concerns can be directed to the Customer Service Manager at 320.229.7057.
- Initiated
- 2013-03-25
- Root cause
- Labeling mix-ups
- Terminated
- 2022-04-22
- Distribution
- Worldwide distribution - US Nationwide and the countries of Canada, South Africa, France, Netherlands, Mauritius, China.
- Recall number
- Z-0641-2022
- Recall status
- Terminated
- CDRH record ID
- 191717
- FDA record created
- 2022-02-14
- Recall event number
- 89479