Open original FDA record ↗Recall record
- FDA identifier
- Z-0643-2009
- Company in FDA record
- Acist Medical Systems
- FDA classification date
- Jan 8, 2009
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
Multiple Injections may result. It has been determined that when using the ACIST CVi Angiographic Injection system to synchronize with the Toshiba Infinix -i cardiovascular x-ray system there is an injection scenario that may result in multiple injections of contract media when one injection was intended.
Source facts
- Action
- ACIST is distributing an Urgent Medical Device Correction letter, dated November 10, 2008, to affected customers. The letter describes the issue and the scenario when the issue to occur, instructs them to discontinue use of the synchronized mode between the ACIST Angiographic Injection System model CVi and the Toshiba Infinix-i X-ray fluoroscopic system by disconnecting the interface cable immediately. The letter states that ACIST is working with Toshiba to resolve this issue. A response form is asked to be returned either by fax, e-mail or mail. If there are any questions, they should be directed to either ACIST at 952-995-9300 or local Sales Reps or Clinical Applications Specialist.
- Initiated
- 2008-11-10
- Root cause
- Software design
- Terminated
- 2011-05-07
- Distribution
- Worldwide Distribution --- USA including states of AR, AZ, FL, and NY and countries of France, Germany, Italy, Japan, and Spain.
- Recall number
- Z-0643-2009
- Recall status
- Terminated
- CDRH record ID
- 75534
- FDA classified
- 2009-01-08
- Recall event number
- 50501