Skip to content

Public FDA record

Z-0644-2013: Laparoscope, general; plastic surgery Used for general, abdominal, gynecological and thoracic minimally

Recall record Z-0644-2013 for Applied Medical Resources Corp. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0644-2013
Company in FDA record
Applied Medical Resources Corp
FDA classification date
Jan 7, 2013
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Applied Medical is conducting a voluntary recall on a specific lot number of its 12x150mm Kii shielded bladed access systems. This is due to a potential for the blade to remain exposed after shield deployment. Further use of any affected product should cease immediately. The affected model number is CTB71, and the affected lot is 1179107.

Source facts

Action
Applied Medical sent an Urgent Medical Device Recall to all affected customers in November 2012. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to check their inventory for the affected product, complete the attached Recall Notification Confirmation Form to acknowledge the recall and indcated any product their facility is returning. The Recall confirmation form should be returned to Applied Medical by emailing to recall60494844#applied medical.com or fax to 949-713-8832. Return affecte4d product and a copy of the confirmation form to Applied Medical. Customers with questions were instructed to call 949-713-8662. For questions regarding this recall 949-713-8662.
Initiated
2012-11-05
Root cause
Nonconforming Material/Component
Terminated
2013-01-16
Distribution
Worldwide Distribution - USA (nationwide) and France.
Recall number
Z-0644-2013
Recall status
Terminated
CDRH record ID
114879
FDA classified
2013-01-07
Recall event number
63842

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗