The attachment of the NuCART Spring Arm to the NuCART horizontal boom may come apart if the securing ring (snap ring) was not installed correctly causing the spring arm to release and detach.
Source facts
Action
The firm began inspecting devices between 12/4/15 and 12/16/15, and sent the Nucart Product Investigation letter to consignees between 12/11/15 and 12/15/15.
Consignees with questions can call the firm at 503-601-5603.
Initiated
2015-12-04
Root cause
Nonconforming Material/Component
Terminated
2016-03-31
Distribution
US nationwide distribution including AZ, CA, CO, GA, OH, and TN.
Recall number
Z-0644-2016
Recall status
Terminated
CDRH record ID
142497
FDA classified
2016-01-13
Recall event number
72915
Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.
Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.