Open original FDA record ↗Recall record
- FDA identifier
- Z-0645-2007
- Company in FDA record
- Siemens Healthcare Diagnostics
- FDA classification date
- Mar 22, 2007
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
Firm received complaints for signal 2 & 4 errors, and shifts in Relative Light Units (RLU). Investigations showed that the On-Board Stability (OBS) of the acid & base reagents may be compromised after 4 days after installation onto the Centaur CP system due to evaporation.
Source facts
- Action
- A Support Bulletin was sent by e-mail to all affected Bayer HealthCare LLC Branches on 11/09/06 in order to inform them of the issue and to provide them with instructions for this field correction. Two Customer Bulletins were prepared and sent by e-mail to Bayer HealthCare LLC Branches on 11/09/06 both in the US and outside the US for communication with the affected customers. The first bulletin advised customers of the problem and the corrective action. A Confirmation (Fax-back) Form was included to ensure that all customers worldwide have been notified and have complied with the mandatory action. The second Customer Bulletin describes an optional alternative mitigation, which requires a Bayer representative to go to the customer site to implement.
- Initiated
- 2006-11-09
- Root cause
- Other
- Terminated
- 2012-07-30
- Distribution
- Worldwide
- Recall number
- Z-0645-2007
- Recall status
- Terminated
- CDRH record ID
- 50464
- FDA classified
- 2007-03-22
- Recall event number
- 37390