Open original FDA record ↗Recall record
- FDA identifier
- Z-0646-2009
- Company in FDA record
- Siemens Medical Solutions USA, Inc
- FDA classification date
- Jan 13, 2009
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
Excessive wear on drive belts used for Image Intensifier and/or X-ray head drive mechanisms may result in uncontrolled movement, which may cause injury.
Source facts
- Action
- Firm has reissued the Urgent Medical Device Customer Letter regarding the belt replacement and a reminder message to replace the belts yearly during the annual Preventative Maintenance. The letter requests that customers follow the guidelines, which consist of making any required SAD adjustments utilizing the Image Intensifier drive, the X-Ray Head Drive, or both; being vigilant of any change in the type or level of noise from either of the drive belt assemblies while they are in use as this may be an indication of wear; discontinuing clinical operation if there is any doubt, and contacting their local Siemens Service organization to inspect the belts. Currently, the firm is distributing the current belt kit to replace the entire installed base of active SIMVIEW Simulators (3000 and NT models).
- Initiated
- 2008-11-06
- Root cause
- Device Design
- Terminated
- 2010-12-09
- Distribution
- Worldwide Distribution
- Recall number
- Z-0646-2009
- Recall status
- Terminated
- CDRH record ID
- 75465
- FDA classified
- 2009-01-13
- Recall event number
- 50480