Open original FDA record ↗Recall record
- FDA identifier
- Z-0646-2013
- Company in FDA record
- Siemens Healthcare Diagnostics
- FDA classification date
- Jan 7, 2013
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class III
Reason stated in the source
The recall was initiated because Siemens Healthcare Diagnostics confirmed an under-recovery of dose values with some vials of Coat-A-Consta (CAC) IRMA TSH calibrators TSI4 (level B), TSI5 (level C), TSI6 (level D), TSI8 (level F), and TSI9 (level G) lot 035. The affected calibrator vials can be identified by a darker yellow color than normal. The TSI3 (level A), TSI7 (level E), and TSIX (level H)
Source facts
- Action
- Siemens sent an Urgent Medical Device Recall letter dated November 2012 to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to discontinue use of the affected product, complete and return the Urgent Medical Device Recall Effectiveness Check form included with the letter by fax to 302-631-7597. Customers were also instructed to determine their replacements needed, and to forward the recall letter to whomever they may have distributed the affected product. For questions customers were instructed to contact their Siemens Technical Solutions Ceter or their local technical support representative. For questions regarding this recalla call 914-524-2955.
- Initiated
- 2012-11-28
- Root cause
- Nonconforming Material/Component
- Terminated
- 2013-10-28
- Distribution
- Worldwide Distribution - USA including CA, GA, IL, IN, MI, MN, NY, and Puerto Rico. Internationally to Agentina, Bolivia, Brazil, Canada, Chile, Guatemala, South Korea, Philippines, Paraguay, Taiwan, and Uraguay,
- Recall number
- Z-0646-2013
- Recall status
- Terminated
- CDRH record ID
- 114905
- FDA classified
- 2013-01-07
- Recall event number
- 63852