- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
Indicated orientation does not match actual orientation of the patient.
Source facts
- Action
- The recalling firm issued a Customer Safety Advisory dated 10/8/08 to affected customers via Update Instructions CT052/08/S. The letter informs customers of the potential issue and provides instructions to avoid its occurrence. Customers should consider "no saving of calculated Dual Energy data sets from non-supine and mirror-option scans via "Save Sets CT"-button." The firm also highly recommends for customers "to check the correct display of "left", "anterior" and "posterior"markings, e.g. by checking the position of the spine with respect to the posterior marking in the image." The error should be fixed within certain upgrade and installation activities on the affected software.
- Initiated
- 2008-11-13
- Root cause
- Software design
- Terminated
- 2013-03-19
- Distribution
- Nationwide Distribution
- Recall number
- Z-0647-2009
- Recall status
- Terminated
- CDRH record ID
- 75531
- FDA classified
- 2009-01-08
- Recall event number
- 50500