Z-0647-2022: Zenition 50, Model #718096

Recall record Z-0647-2022 for PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0647-2022
Company in FDA record
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
FDA record creation date
Feb 17, 2022
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Firm discovered that the wireless foot switch can suddenly stop responding due to a firmware issue. If the wireless foot switch fails, there could be a delay or interruption of procedure if the hand switch is not used or if the wired foot switch is not connected to the system.

Source facts

Action
On December 2, 2021, the firm issued Urgent Medical Device Correction letters to affected customers. The Medical Device Correction includes the following instructions for the Customer: o Ensure that the wired foot switch is always connected to the system. o Immediately start using the wired foot switch or hand switch in the event the wireless foot switch loses connection. o Perform a daily check for damage and correct functioning of the hand switch and wired foot switch as described by the User Routine Checks section in your Instruction for Use. o Circulate this notice to all users of this device so that they are aware of the issue./ o Place this Medical Device Correction with the documentation of the Philips Zenition 50, Zenition 70 or Veradius Unity system. o Return the attached reply form to Philips to confirm that: 1. The Zenition 50, Zenition 70 and Veradius Unity systems wired foot switch is connected to the system 2. The users of the system have reviewed and understood this Medical Device Correction. If you need further information or support concerning this issue, please contact your local Philips representative: 1-800-722-9377.
Initiated
2021-12-02
Root cause
Under Investigation by firm
Distribution
Domestic distribution to AZ CA FL GA HI IL KS KY MA MD ME MI MN MS NC NE NJ NV NY OH PA TN TX VA WA WV. Worldwide foreign distribution.
Recall number
Z-0647-2022
Recall status
Open, Classified
CDRH record ID
191569
FDA record created
2022-02-17
Recall event number
89437

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗