Open original FDA record ↗Recall record
- FDA identifier
- Z-0648-2012
- Company in FDA record
- Navilyst Medical, Inc
- FDA classification date
- Jan 11, 2012
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 17, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
Navilyst Medical, Inc. determined that the Exodus Standard Loop Biliary Drainage Catheters exhibited a rate of reported failures (catheter tip fragmentation) higher than anticipated.
Source facts
- Action
- The firm, Navilyst Medical, sent an "Urgent Medical Device Recall -Immediate Action Required" letter dated February 4, 2011, via Federal Express to the domestic consignees/customers and one foreign consignee/customer. Consignees/customers were instructed to: (1.) Immediately discontinue use and remove the recalled product from inventory; (2.) Segregate this product in a secure location for return to Navilyst Medical, Inc.; (3.) Immediately forward a copy of the recall notification to ALL sites to which they have distributed affected device (4.) promptly complete and return the Reply Verification Tracking Form (even if you do not have any product to return) via fax to: QA Manager, Navilyst Medical, Inc., 1-800-782-1357; (5.) call Customer Service at 1-800-833-9973 between 8:30.am to 7:00pm EST to obtain RMA number; (6.) package and return the recalled product; and (7.) Physicians were instructed to continue to monitor their patients and evaluate per institutional protocol for biliary catheters. If you have any questions about this recall action, your local Sales Representative is available to assist you or contact QA Manager, Navilyst Medical, Inc., directly at 518-742-4571.
- Initiated
- 2011-02-04
- Root cause
- Nonconforming Material/Component
- Terminated
- 2012-01-10
- Distribution
- Worldwide distribution: USA (nationwide) including states of: AK, AZ, CA, CO, FL, IL, KS, MI, MO, NC, NJ, NY, OH, OK, PA, PR and country of: South Africa.
- Recall number
- Z-0648-2012
- Recall status
- Terminated
- CDRH record ID
- 106450
- FDA classified
- 2012-01-11
- Recall event number
- 60775