The product is not uniform across the cassettes of some lots, which will result in an inconsistent bias of patient results.
Source facts
Action
Recall letters dated 2/15/05 were sent to each user instructing customers to run controls before the use of each cassette or kit in a lot, instead of once per lot as previously instructed, and to notify the physician at the facility.
Initiated
2005-02-15
Root cause
Other
Terminated
2005-05-17
Distribution
Nationwide.
Recall number
Z-0649-05
Recall status
Terminated
CDRH record ID
37073
FDA classified
2005-03-23
Recall event number
31056