Handle could come off. The bolts may back out over time, resulting in a risk of dropping the patient and/or injury to transport personnel, a delay in medical treatment, or a delay in evacuation.
Source facts
Action
Consignees were notified by Urgent Medical Device Correction letter dated 11/18/08 and given the serial numbers of their units, an upgrade kit and instructions for performing replacement of the bolts. Consignees were informed Stryker will perform the upgrade if requested.
Initiated
2008-11-18
Root cause
Process control
Terminated
2010-10-27
Distribution
Worldwide Distribution: USA, India, Netherlands and United Kingdom.
Recall number
Z-0650-2009
Recall status
Terminated
CDRH record ID
75006
FDA classified
2009-02-04
Recall event number
50178