Revised Battery Instructions for Removal and Expiration Date due to battery swelling and cell leakage
Source facts
Action
On November 18, 2013, an URGENT Field Correction Letter was sent to all StatStrip Glucose Hospital meter, StatStrip Lactate Hospital Meter and StatSensor Creatinine Hospital Meter customers via email to the Point of Care Coordinator to alert users that lithium batteries that are utilized beyond their intended useful life may swell and/or crack posing a potential safety hazard. In some cases, the swelling may potentially cause the battery door on the meter to dislodge and render the meter inoperable.
Initiated
2013-10-10
Root cause
Component design/selection
Terminated
2014-03-24
Distribution
Nationwide Distribution.
Recall number
Z-0650-2014
Recall status
Terminated
CDRH record ID
122250
FDA classified
2014-01-09
Recall event number
66398