Skip to content

Public FDA record

Z-0650-2022: Merete PediatrOS RigidTack, REF CP20120, Size 20 mm, Sterile, Rx Only

Recall record Z-0650-2022 for Merete Medical GmbH. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0650-2022
Company in FDA record
Merete Medical GmbH
FDA record creation date
Feb 17, 2022
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Incorrect marketing label (25 mm) was applied to product size 20 mm.

Source facts

Action
Customers were notified of the recall via phone call on 01/19/2022 and instructed to quarantine any devices from the recalled lot. The recalling firm also informed customers that a written notice was going to be distributed with further instructions. Devices remaining in stock should be separated and quarantined, while those implanted should have patients informed and x-rays conducted to evaluate if the appropriate size was utilized. Additionally, customers were notified via email on 01/27/2022 of the recall in an Urgent Medical Device Recall Notice. The notice identified Lot MS2117816 of PediatrOS RigidTrack staples (Model No. CP20120) as being recalled due to mislabeling. Customers were asked to identify any recalled product in stock, quarantine identified units, and complete the provided Recall Response Form to facilitate the return of affected devices. If product has been implanted already, the firm asks that surgeons verify that the staple is adequately positioned surrounding the epiphyseal cartilage. Completed response forms are to be faxed to +49 (0)30 77 99 80 - 177 or emailed to safety@merete.de. Customers with any questions relating to the recall are encouraged to contact Mr. Paul Munch, Safety Officer, as +49 30 77 99 80 - 181 from 2:00 AM to 10:00 AM Eastern Times Zone, or by phone at +49 151 1881 6304 outside of office hours. He is also available by email at safety@merete.de.
Initiated
2022-01-18
Root cause
Error in labeling
Terminated
2023-02-16
Distribution
Domestic: Illinois; Foreign: Germany.
Recall number
Z-0650-2022
Recall status
Terminated
CDRH record ID
191721
FDA record created
2022-02-17
Recall event number
89482

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗