Open original FDA record ↗Recall record
- FDA identifier
- Z-0651-2016
- Company in FDA record
- Midmark Corp dba Progeny Inc
- FDA record creation date
- Jan 25, 2016
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Failure of the primary collimator may result in unintentional movement of the device. The unintentional movement will not result in the generation of an error message.
Source facts
- Action
- MidMark sent an Urgent Medical Device Field Correction letter dated September 4, 2014, to all affected customers. The letter included instructions for distributors to: 1) encourage end users to contact Midmark to arrange the field correction after the final solution becomes available; 2) if a partial image is reported to have occurred with a machine, the machine should be removed from use and the end user should initiate a service request per the normal procedure; 3) quarantine any affected machines in the distributor's inventory and contact Midmark Technical Support within 10 business days to provide the Serial numbers to on-site servicing of the affected machines can be scheduled. Distributors with questions about this market action can contact Midmark Technical Support by calling 888-294-3800 x 3 or by e-mail at imagingtechnsupport@midmark.com.
- Initiated
- 2014-09-04
- Root cause
- Device Design
- Terminated
- 2017-12-29
- Distribution
- Worldwide Distribution - US including AL, AR, AZ, CO, CT, DE, FL, GA, IA, IL, IN, KS, KY, LA, MA, MI, MN, MO, MS, NE, NJ, NM, NV, NY, OH, OK, PA, SC, TN, TX, UT, VA, WA, WI; and Internationally to Australia, Belgium, Canada, Czech Republic, France, Hungary, Lithuania, New Zealand, Poland, Russia, Saudi Arabia, South Africa, Switzerland, Turkey, United Arab Emirates, United Kingdom, and Vietnam.
- Recall number
- Z-0651-2016
- Recall status
- Terminated
- CDRH record ID
- 134633
- FDA record created
- 2016-01-25
- Recall event number
- 70765