Skip to content

Public FDA record

Z-0652-03: Optima URS X-Ray System; Manufactured by SEDECAL, C/Pelya 9 Pol. Ind. Rio De Janeiro, 28110 Algete (Madri

Recall record Z-0652-03 for Sedecal USA, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0652-03
Company in FDA record
Sedecal USA, Inc.
FDA classification date
Apr 25, 2003
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

Incorrect identification labels and their dertification testing program faialed in noncompliance with U.S. Federal performance standards.

Source facts

Action
Sedecal USA sent end user notification letters dated 9/24/02 to the consignees having the affected Sedecal Stationary X-Ray Systems, informing them that the units do not meet with the U.S. federal performance standards, that the labels are in Spanish and that the labels may not be visible after system installation. Sedecal schedule an appointment to conduct performance tests on the mobile x-ray systems to verify their compliance with the federal performance standard, submit the data to FDA, and place the English labels on the units. Sedecal will reimburse the owners for any out-of-pocket expenses incurred. Sedecal requested the accounts to call 800-920-9525 if they had any questions, and to inform Sedecal if they sold the unit, destroyed the unit or otherwise no longer had the unit in their possession.
Initiated
2002-09-24
Root cause
Other
Terminated
2005-07-29
Distribution
New York, Texas, Florida and Minnesota.
Recall number
Z-0652-03
Recall status
Terminated
CDRH record ID
27098
FDA classified
2003-04-25
Recall event number
26111

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗