Z-0652-2025: ChemoLock Close Vial Spike REF CL-80 The Chemolock Closed System Transfer Device prevents the transfer o

Recall record Z-0652-2025 for ICU Medical, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0652-2025
Company in FDA record
ICU Medical, Inc.
FDA classification date
Dec 9, 2024
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Due to finished goods incorrectly packaged and labeled.

Source facts

Action
On 11/06/2024, the firm send an "URGENT Medical Device Correction" Letter via third party (Sedgwick, Inc.) to customers informing them that product was incorrectly packaged and labeled. Product CL-80 package incorrectly contained product code CL-80S. The CL-80 is a used with a 20 mm or a 28 mm vial where as CL-80S is only use for a 20 mm vial. The issue is using the incorrect adapter may result in leakages. Actions to be taken by the Customer: Discontinue use mislabeled lot 13953938 of CL-80 and discard remaining product from this specific lot. 1. Check all inventory locations within your institution for the affected product listed in the notification and discontinue use. Quarantine and destroy all affected products following your institution s process for destruction. If destroying is not immediately possible at your facility, then the product should be quarantined until disposal. 2. Share this notification with all potential users of the device to ensure they are aware of this notification. If the devices are used at another location, please ensure this communication is delivered there./// 3. Complete and return the attached Customer Response Form to XXXX@sedgwick.com within 10 days of receipt to acknowledge your understanding of this notification./ / 4. DISTRIBUTORS: If you have distributed affected products to your customers, please immediately forward this notice to them. Request that they complete the response form and return it to XXXX@sedgwick.com. For questions/assistance, contact: Global Complaint Management 1-844-654-7780 (M-F, 8:00am 5:00pm CT) or ProductComplaintsPP@icumed.com To report adverse events or product complaints Customer Service 1-866-829-9025, option 8 or customerservice@icumed.com (M-F, 8:00 am 6:00 pm CT) Additional information or to request a credit. Field Action Processing 1-888-843-0245 xxx@sedgwick.com Questions about this communication
Initiated
2024-11-06
Root cause
Packaging process control
Distribution
Worldwide - U.S. Nationwide distribution in the state of IL and the country of Canada.
Recall number
Z-0652-2025
Recall status
Open, Classified
CDRH record ID
210887
FDA classified
2024-12-09
Recall event number
95649

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗