Open original FDA record ↗Recall record
- FDA identifier
- Z-0652-2025
- Company in FDA record
- ICU Medical, Inc.
- FDA classification date
- Dec 9, 2024
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Due to finished goods incorrectly packaged and labeled.
Source facts
- Action
- On 11/06/2024, the firm send an "URGENT Medical Device Correction" Letter via third party (Sedgwick, Inc.) to customers informing them that product was incorrectly packaged and labeled. Product CL-80 package incorrectly contained product code CL-80S. The CL-80 is a used with a 20 mm or a 28 mm vial where as CL-80S is only use for a 20 mm vial. The issue is using the incorrect adapter may result in leakages. Actions to be taken by the Customer: Discontinue use mislabeled lot 13953938 of CL-80 and discard remaining product from this specific lot. 1. Check all inventory locations within your institution for the affected product listed in the notification and discontinue use. Quarantine and destroy all affected products following your institution s process for destruction. If destroying is not immediately possible at your facility, then the product should be quarantined until disposal. 2. Share this notification with all potential users of the device to ensure they are aware of this notification. If the devices are used at another location, please ensure this communication is delivered there./// 3. Complete and return the attached Customer Response Form to XXXX@sedgwick.com within 10 days of receipt to acknowledge your understanding of this notification./ / 4. DISTRIBUTORS: If you have distributed affected products to your customers, please immediately forward this notice to them. Request that they complete the response form and return it to XXXX@sedgwick.com. For questions/assistance, contact: Global Complaint Management 1-844-654-7780 (M-F, 8:00am 5:00pm CT) or ProductComplaintsPP@icumed.com To report adverse events or product complaints Customer Service 1-866-829-9025, option 8 or customerservice@icumed.com (M-F, 8:00 am 6:00 pm CT) Additional information or to request a credit. Field Action Processing 1-888-843-0245 xxx@sedgwick.com Questions about this communication
- Initiated
- 2024-11-06
- Root cause
- Packaging process control
- Distribution
- Worldwide - U.S. Nationwide distribution in the state of IL and the country of Canada.
- Recall number
- Z-0652-2025
- Recall status
- Open, Classified
- CDRH record ID
- 210887
- FDA classified
- 2024-12-09
- Recall event number
- 95649