Skip to content

Public FDA record

Z-0653-2014: Product Brand Name(s): Innovacon Oral Fluid Drug Screen Device, OrALert Oral Fluid Drug Screen Device,

Recall record Z-0653-2014 for Alere San Diego, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0653-2014
Company in FDA record
Alere San Diego, Inc.
FDA classification date
Jan 9, 2014
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Alere initiated this recall of certain lots of the Oral Fluid Drug Screen Device because the sponge may become dislodged from the Saliva Collector handle prior to or during the oral fluid sample collection process. If the sponge becomes dislodged during the specimen collection process, the donor may inadvertently swallow the sponge which could present a choking risk.

Source facts

Action
Firm sent customers notification letter via combination of fax, email or direct mail. The notification letter dated June 3, 2013, titled "URGENT PRODUCT SAFETY NOTICE", informed customers of the recall, reason for recall, product description with lot codes, potential hazard statement, customer required actions, and contact information.
Initiated
2013-06-03
Root cause
Other
Terminated
2014-03-18
Distribution
US Distribution: including states of: AL, CA, PA, and VA.
Recall number
Z-0653-2014
Recall status
Terminated
CDRH record ID
123825
FDA classified
2014-01-09
Recall event number
66870

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗