Skip to content

Public FDA record

Z-0653-2022: Medfusion Model 3500 and 4000 Syringe Infusion Pumps and Medfusion 3010 and 3010A Syringe Infusion Pumps

Recall record Z-0653-2022 for Smiths Medical ASD Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0653-2022
Company in FDA record
Smiths Medical ASD Inc.
FDA classification date
Feb 17, 2022
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

There have been intermittent issues with Supercap POST (power on self-test) and Backup Audible Alarm error codes outside of the pump warranty period. These alarms occur at start up when the pump goes through the self-test. This recall occurred in 2015.

Source facts

Action
The firm issued Customer Information Bulletin dated 03-Dec-2015. It explained the details of the issue and provides the consignee a list of serial numbers for pumps in their facility that may be impacted. In addition, it informs the consignee if they have purchased main boards between April 2012 and June 2015 for part numbers G6000361, G6000435, G6001260, G6001560 and G6002728, they are included in the enclosed file. The consignee is to contact the recalling firm to make arrangements to begin the replacement process. If the Supercap Post alarm or the Backup Audible Alarm error code occurs prior to replacement, the consignee is to immediately remove the pump from service for repair by a trained biomedical technical at the recalling firm or at their facility. If the device with the affected serial number is sent to the recalling firm for any other unrelated repair, the supercapacitor will be replaced at that time. No response form was enclosed with the bulletin.
Initiated
2015-12-03
Root cause
Component change control
Distribution
Distribution US nationwide, including PR, VI, GU, and MP. There was government/military/foreign distribution.
Recall number
Z-0653-2022
Recall status
Open, Classified
CDRH record ID
190636
FDA classified
2022-02-17
Recall event number
89144

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗