Z-0654-2021: Star 8 Rolling Walker/Rollator, Items: 4258/ BK(Black)/ BL(Blue)/ DB(Sky Blue)/ PL(Purple)/ RD(Red)

Recall record Z-0654-2021 for Nova Ortho-Med Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0654-2021
Company in FDA record
Nova Ortho-Med Inc
FDA record creation date
Dec 21, 2020
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Walkers within the identified serial range have the potential for the front fork to unscrew itself from the leg. A partially unscrewed fork creates unevenness in the wheelbase, which could cause instability, that if not detected by the user, may result in the front fork falling off the walker.

Source facts

Action
On 10/07/2019, Product Correction and Product Safety Recall notices were emailed to customers. Customers were asked to replace the front leg assemblies as follows: a. Depress the push buttons on the original legs and remove them from the frame. b. Insert the improved replacement assemblies issued by NOVA into the unit and discard the original legs. In addition, customers were asked to do the following: 1. Check your inventory for affected walkers and fix. 2. Contact all customers who purchased affected walkers notify them of the potential issue. Schedule a time that they can come in for repair or if the customer prefers the recalling firm can send the wheel assemblies directly to their residence. Have them inspect their walker immediately to identify if either of the front forks have unscrewed from the leg. If so, please instruct them to not use the walker until the leg assemblies can be replaced with the new parts. 3. Call the recalling firm to request the wheel assembly set to be sent to the required location. 4. Record the serial #'s of the product that have been corrected and email them to Customer Service (info@novajoy.com) so that we can have for our records. If you prefer to call, then dial 1-800-557-6682 x8129. Email: info@novajoy.com
Initiated
2019-10-07
Root cause
Device Design
Distribution
U.S.: AK, AR, AZ, CA, CO, CT, FL, GA, HI, IA, ID, IL, IN, KS, KY, MA, MD, MI, MN, MO, MT, NC, NE, NJ, NM, NV, OH, OK, OR, PA, SC, SD, TN, TX, UT, WA, WI, WY O.U.S.: Taiwan
Recall number
Z-0654-2021
Recall status
Open, Classified
CDRH record ID
184874
FDA record created
2020-12-21
Recall event number
86941

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗