Open original FDA record ↗Recall record
- FDA identifier
- Z-0655-2016
- Company in FDA record
- Skytron, Div. The KMW Group, Inc
- FDA classification date
- Jan 14, 2016
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
The hardware that holds the VESA plate and the display interface / monitor to the bracket may fail causing the display interface to fall from the bracket. Over time and with use the threaded rod hardware material can weaken and detach from the bracket. A display interface falling from its mounting location may result in injury to the patient or provider.
Source facts
- Action
- Skytron sent an Urgent Medical Device Recall letter dated December 02. 2015, to all affected customers via certified mail. The recall notification included a description of the reason for the recall, affected product, consignee responsibilities, and instructions for responding to the formal recall notification. Customers with questions were instructed to contact at 616-656-2900 with any questions. For after-hours support, please call 1-800-SKYTRON.
- Initiated
- 2015-12-02
- Root cause
- Device Design
- Terminated
- 2016-07-19
- Distribution
- Worldwide Distribution - US including WI, TX, FL, NC, MD, OH, AZ, CA, UT, MT, OK, HI, CO, WY, MO, NJ, IL, NY, MA, IA, DE, PA, MI, IN, and WV and Internationally to Canada, and Japan.
- Recall number
- Z-0655-2016
- Recall status
- Terminated
- CDRH record ID
- 142447
- FDA classified
- 2016-01-14
- Recall event number
- 72897