Z-0655-2022: Digital Angiography System Shimadzu Trinias Generator Model: D150GC-40

Recall record Z-0655-2022 for Shimadzu Medical Systems. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0655-2022
Company in FDA record
Shimadzu Medical Systems
FDA record creation date
Feb 28, 2022
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Due to the inadequate adjusting criteria in installation, the system generator on two Digital Angiography Systems, the irradiated x-ray may exceed the x-ray radiation dose rate in fluoroscopic mode.

Source facts

Action
On 12/13/2021, the firm sent an "Urgent: Voluntary Medical Device Recall Notice" to distributors who hand-delivered to customers informing them that there is a potential issue with the adjustment procedure for the calibration of the x-ray generators for the Digital Angiography Systems and occurs during the calibration of the fluoroscopic mode which could cause x-ray radiation dose rate to exceed regulation and/or standards. Customers are instructed to: -Check their systems for the affected generators. -Contact their distributor to schedule a system update and calibration. -If customers feel there are any issues with their system, they are asked to discontinue use of the system and contact their local service providers. The Recalling Firm has identified that affected units and have informed the authorized service representatives. The authorized servicers should be contacting customers to schedule time to implement the corrective action (update and calibrate). Customers can contact the Recalling Firm's National Technical Support at: Toll Free: (844)487-2767 available 8:00 a.m. to 5:00 p.m. Monday thru Friday On 01/07/2022, an updated "Urgent: Voluntary Medical Device Recall Letter" will be communicated to customer to provide additional instruction on how to identify an issue with their system which would indicated to cease using until the system has been corrected.
Initiated
2021-12-16
Root cause
Process change control
Distribution
U.S.: CA, FL, GA, IL, IN, KY, LA, MI, MS, OH, PA, SC, TN, TX and WA O.U.S.: Not provided.
Recall number
Z-0655-2022
Recall status
Open, Classified
CDRH record ID
190935
FDA record created
2022-02-28
Recall event number
89253

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗