When the user changes frame rates from lower frequency (e.g. 10 f/s) to higher frequency (e.g. 30 f/s) during continuous release of Fluoro "Fluoro Override", the measured Air Kerma Rate can exceed the regulatory limits.
Source facts
Action
Siemens Medical Solutions USA, Inc. has initiated a Customer Advisory Notice to all affected customers via AX067/20/P on November 30, 2020. All modifications and repairs are provided to affected customers free of charge.
Initiated
2020-11-30
Root cause
Software design
Distribution
Worldwide-US Nationwide
Recall number
Z-0656-2021
Recall status
Open, Classified
CDRH record ID
185060
FDA record created
2021-01-11
Recall event number
87026