Open original FDA record ↗Recall record
- FDA identifier
- Z-0657-2013
- Company in FDA record
- Stryker Communications, Inc.
- FDA classification date
- Jan 9, 2013
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Stryker has become aware that there is a low likelihood of incomplete engagement of the Spring Arm Circlip during service or replacement of the Spring Arm. If the Spring Arm Circlip is not fully seated, the Spring Arm could separate and fall along with the surgical lights or monitors mounted to the end of the Spring Arm. To date, there have been no serious injuries reported from the incomplete e
Source facts
- Action
- Stryker sent an Urgent Medical Device Correction letter dated September 25, 2012, to all affected customers. The letter identified the product, the prolem, and the action to be taken by the customer. Customers were instructed to immediately inspect their Spring Arm. If customers experienced any concerns with their Spring Arm they were instructed to contact Stryker immediately at 1-866-841-5663 to schedule a service technician to inspect and correct the Spring Arm circlips. For questions regarding this recall call 972-410-7100.
- Initiated
- 2012-09-25
- Root cause
- Equipment maintenance
- Terminated
- 2013-04-22
- Distribution
- Worldwide Distribution - USA (nationwide) and Internationally to Australia, Japan, Taiwan, Canada, United Kingdom, France, Netherlands, Spain, Mexico, India, Korea, Singapore, New Zealand, Hong Kong, Philippines and Brazil.
- Recall number
- Z-0657-2013
- Recall status
- Terminated
- CDRH record ID
- 113562
- FDA classified
- 2013-01-09
- Recall event number
- 63355