Open original FDA record ↗Recall record
- FDA identifier
- Z-0657-2019
- Company in FDA record
- Howmedica Osteonics Corp.
- FDA record creation date
- Dec 21, 2018
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Instead of having a spherical diameter, the probe was observed to have 2 radii and a flat surface. Additionally, upon further investigation, it was observed that some parts exhibited an edge on the intersection between the radius and the diameter of the probe.
Source facts
- Action
- The firm, Stryker, sent an "Urgent Medical Device Recall" letter dated 11/14/2018 to its customers on 11/26/2018. The letter described product, problem and actions to be taken. The customers were instructed to: Examine your inventory; Immediately quarantine the affected product; and complete customer response form to Stryker Spine at email: Spine-RegulatoryAction@stryker.com. For questions contact Manager-Regulatory Compliance & Quality System at 866-45-SPINE (77463)..
- Initiated
- 2018-11-26
- Root cause
- Under Investigation by firm
- Terminated
- 2020-05-26
- Distribution
- Worldwide distribution: US Distribution to the state of: PA and countries of:: Australia, Canada, France, Italy, Netherlands, Poland, and Spain.
- Recall number
- Z-0657-2019
- Recall status
- Terminated
- CDRH record ID
- 169205
- FDA record created
- 2018-12-21
- Recall event number
- 81645