Skip to content

Public FDA record

Z-0658-2022: NobelReplace Conical Connection RP 4.3x11.5 mm REF 36707 STERILE R

Recall record Z-0658-2022 for Nobel Biocare Usa Llc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0658-2022
Company in FDA record
Nobel Biocare Usa Llc
FDA record creation date
Feb 18, 2022
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Due to 16 complaints related to drivers not engaging in implants. According to the firm's investigations, the incorrect offset gauge was potentially used during production of the implants. As a result, this could potentially lead to aborted or a delay in surgical procedure.

Source facts

Action
On 01/12/2022, the firm sent an "URGENT MEDICAL DEVICE RECALL" Letter via UPS to customers informing them that specific batches/lots of devices, the dimensions of the hexagon connection part are potentially not fully conforming to specification and may not fully engage in the implant. Customers are instructed to: 1. Inspect their stock and quarantine all affected devices. 2. Complete the Customer Acknowledgement Form and return to the Recalling Firm via email or fax within 5 days of receipt of this notice. 3. Return all affected stock on hand to the Recalling Firm using the shipping label provided. 4. Ensure relevant staff members are informed of this recall, and if you further supplied or transferred any of the affected products to another facility or organization, let these facilities know of the recall and provide them with a copy of this notice. For questions, contact your local customer support representative,
Initiated
2021-12-17
Root cause
Under Investigation by firm
Terminated
2024-10-25
Distribution
Worldwide distribution - U.S. Nationwide distribution in the states of AL, AR, AZ, CA, CO, FL, GA, HI, ID, IL, IN, KS, KY, MA, ME, MI, MN, MS, NC, NJ, NM, NV, NY, OH, OR, PA, SC, TN, TX, UT, VA, VT, WA and WI. The countries of Canada, Mexico and Turkey.
Recall number
Z-0658-2022
Recall status
Terminated
CDRH record ID
191694
FDA record created
2022-02-18
Recall event number
89465

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗