Open original FDA record ↗Recall record
- FDA identifier
- Z-0659-2016
- Company in FDA record
- Cyberonics, Inc
- FDA classification date
- Jan 15, 2016
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Recall being initiated in response to three reports of "Burst Watchdog Timeout" events occurring with the Model 106 AspireSR Generator, resulting in a device reset condition where stimulation output is disabled.
Source facts
- Action
- The firm notified consignees of the issue via letter on 12/18/15. The letter identified the affected device, the issue involved, and actions to be taken. Physicians are to contact Clinical Technical Support at 866-882-8804 to report if a patient's generator has been disabled due to the issue identified. Users are to complete and return the effectiveness card as soon as possible. If further information is needed, customers can contact Clinical Technical Support at 866-882-8804 or via e-mail at cservices@livanova.com.
- Initiated
- 2015-12-18
- Root cause
- Device Design
- Terminated
- 2017-03-02
- Distribution
- Worldwide Distribution -- United States, Austria, Belgium, Croatia, Czech Republic, Finland, France, Germany, Iceland, Italy, Netherlands, Norway, Poland, Slovakia, Spain, Sweden, Switzerland, United Kingdom, Cyprus, Israel, Kuwait, Lebanon, Oman, Qatar, Saudi Arabia, and United Arab Emirates.
- Recall number
- Z-0659-2016
- Recall status
- Terminated
- CDRH record ID
- 142446
- FDA classified
- 2016-01-15
- Recall event number
- 72896