Z-0659-2017: 5000 Series Freedom Drivers. Freedom Driver System for Temporary Total Artificial Heart (TAH-t). Part n

Recall record Z-0659-2017 for SynCardia Systems Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0659-2017
Company in FDA record
SynCardia Systems Inc.
FDA record creation date
Dec 8, 2016
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class I

Reason stated in the source

The Main Printed Circuit Board Assembly (PCBA) of the affected Freedom Drivers may fail and cause the Freedom Driver to stop functioning without visual or audible alarms, resulting in the loss of life-sustaining function.

Source facts

Action
On 10/21/16 customers were notified by phone. Customers were informed that patients who were being supported by affected Drivers must be switched to unaffected Drivers. The firm will provide unaffected Freedom Drivers to replace the affected Drivers. Affected Drivers must be return to the firm. Instructions for return provided, and Acknowledgment Form must be completed. Contact Information For Questions Mary Pat Sloan Senior Vice President Global Certification and Logistics SynCardia Systems, LLC 1992 E Silverlake Rd Tucson, AZ 85713 Direct 520-547-7470 Toll-free 866-480-1122 Ext 1308 Monday-Friday 7:30 AM to 5:00 PM Pacific/Arizona time Hotline 24/7 866-771-9437 www.syncardia.com On 10/27/16 customer notification letters were sent. The letter provided instructions on how to recognize the device could fail. The patient and caregiver will recognize that an affected Driver is failing because the Freedom Driver will stop pumping without generating any sound or any visual or audible alarms. When functioning, the Freedom Driver exhibits a distinctive pumping sound; when it stops pumping the lack of sound is noticeable. Generally, if the Freedom Driver function is interrupted, a visual and/or audible alarm will occur. However, as a result of this component failure, when affected Freedom Drivers stop functioning, there will be no visual or audible alarms, and no pumping sound. Actions to Be Taken by The Customer/User SynCardia has provided unaffected Freedom Drivers to the two hospitals to replace the affected 5000 series Drivers. All patients with affected Freedom Drivers have undergone the exchange and received unaffected Drivers. The hospitals have returned all the affected Freedom Drivers to SynCardia. The hospitals are requested to complete the Recall Acknowledgement and Receipt Form below and return to SynCardia.
Initiated
2016-10-21
Root cause
Process control
Terminated
2017-02-21
Distribution
US distribution to Virginia and Arizona.
Recall number
Z-0659-2017
Recall status
Terminated
CDRH record ID
150882
FDA record created
2016-12-08
Recall event number
75600

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗