St Jude Medical Cardiac Rhythm Management Division
FDA classification date
Jan 21, 2016
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class I
Reason stated in the source
Leads may have been compromised during the manufacturing process, with cuts in the insulation layer of the SVC shock coil.
Source facts
Action
An Important Medical Device Advisory letter was sent to doctors on 11/3/15 to inform them of the recall by St. Jude Medical. The letter informs the doctors that St. Jude Medical identified that during the manufacturing process of a limited number of Optisure leads, a trim technique to remove excess medical adhesive around the SVC shock coil may have introduced damage to the leads insulation. The letter provides the doctors with the recommendations for patients with a potentially impacted Optisure lead connected to a device with Dynamic TX technology and for patients implanted with a potentially impacted Optisure lead connected to a device without DynamicTx technology. Doctors with questions about the advisory are instructed to contact their local Sales Representative or St. Jude Medical Technical Services at (800) 722-3774.
A revised dear doctor letter was sent via FedEx on 1/21/16 to all affected physicians.
A press release will be sent/posted on 1/22/16.
Initiated
2015-11-03
Root cause
Process control
Terminated
2019-05-10
Distribution
Worldwide Distribution US (nationwide), Belgium, Switzerland, Germany, France, United Kingdom, India, Italy, Japan, Luxembourg, The Netherlands, Saudi Arabia, and Sweden.
Recall number
Z-0660-2016
Recall status
Terminated
CDRH record ID
141537
FDA classified
2016-01-21
Recall event number
72569
Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.