Open original FDA record ↗Recall record
- FDA identifier
- Z-0660-2022
- Company in FDA record
- BALT USA, LLC
- FDA record creation date
- Feb 22, 2022
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Due to Label/Pouch mix-up. Pouches incorrectly labeled. Incorrect shelf carton label or pouch label due to inadequate manufacturing line clearance. As a result of this labeling mix-up, there is the possibility of inadvertently selecting the incorrect coil size for the procedure potentially causing some harms like vessel damage, an inability to treat with said device and potential rupture of the aneurysm.
Source facts
- Action
- On 01/04/2022, the firm sent an "URGENT FIELD SAFETY NOTICE (FSN)/PRODUCT RECALL" letter via email to customers notifying them of the labeling mix-up which presents a high risk to patient safety and product performance, with the possibility that the user could inadvertently select the wrong size coil for a procedure. Customer/Distributors are instruction to: -Inform their customers about the notice; -Identify and locate the affected products -Collect and put in quarantine the affected products and return them to the Recalling Firm through the Return Material Authorization procedure by contacting the Recalling Firm's Customer Service Department. -Inform the Recalling Firm about the status of every unit of the affected products. -Fulfill the Notice of receipt then return to the Recall Firm via the indicated contact For Hospital Staff: -Within their hospital, inform all Safety Officers, Pharmacists, Head of Neuroradiology and the Neuroradiology Department Staff as well as any other person if deemed necessary. -Identify and locate all affected products and cease or stop using. -Collect and put in quarantine all affected products and return them to their local distributor as per its return procedure; -Inform their local distributor and the Recalling Firm about the status of every unit of affected product. -Contact their local distributor or the Recalling Firm for any additional informaiton. For questions or assistance, contact Quality Department at: email QA@balt-usa.com address: 29 Parker Suite 100, Irvine CA 92618, USA Telephone number: 949-788-1443 or Fax 949-788-1444
- Initiated
- 2022-01-04
- Root cause
- Process control
- Distribution
- Worldwide distribution - U.S. Nationwide distribution in the states of FL and MD. The countries of Denmark, France, Germany, Lebanon, Netherlands, Sweden, Peru, Portugal, Slovenia, Spain, and United Arab Emirates.
- Recall number
- Z-0660-2022
- Recall status
- Open, Classified
- CDRH record ID
- 191293
- FDA record created
- 2022-02-22
- Recall event number
- 89338