Reason stated in the source

Internal testing revealed that some of the sensors might not meet the accuracy specifications of : 2% for SpO2 and of: 3% for SpCO, due to the tolerance of one of the emitter LED's used in the calculation ofSpO2 and SpCO. The SpO2 reading may have a negative bias of up to 1 % and the SpCO reading may have a negative bias of up to 4%.

Source facts

Action
Masimo notified all customers by letter sent via Federal Express who were shipped Rainbow DC- I-DC sensors to return them to Masimo for repair or replacement. A Tracking/Verification form will be mailed with the letter to be faxed back to Masimo. Masimo will repair or replace the returned units and return them back to the customer.
Initiated
2006-02-02
Root cause
Other
Terminated
2008-06-17
Distribution
Nationwide and Belgium, Canada, France, Germany, Italy, South Korea, new Zealand, Slovenia, Spain, Sweden, Switzerland, and United Arab Emirates
Recall number
Z-0661-06
Recall status
Terminated
CDRH record ID
44116
FDA classified
2006-03-18
Recall event number
34569

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗