Open original FDA record ↗Recall record
- FDA identifier
- Z-0661-06
- Company in FDA record
- Masimo Corporation
- FDA classification date
- Mar 18, 2006
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
Internal testing revealed that some of the sensors might not meet the accuracy specifications of : 2% for SpO2 and of: 3% for SpCO, due to the tolerance of one of the emitter LED's used in the calculation ofSpO2 and SpCO. The SpO2 reading may have a negative bias of up to 1 % and the SpCO reading may have a negative bias of up to 4%.
Source facts
- Action
- Masimo notified all customers by letter sent via Federal Express who were shipped Rainbow DC- I-DC sensors to return them to Masimo for repair or replacement. A Tracking/Verification form will be mailed with the letter to be faxed back to Masimo. Masimo will repair or replace the returned units and return them back to the customer.
- Initiated
- 2006-02-02
- Root cause
- Other
- Terminated
- 2008-06-17
- Distribution
- Nationwide and Belgium, Canada, France, Germany, Italy, South Korea, new Zealand, Slovenia, Spain, Sweden, Switzerland, and United Arab Emirates
- Recall number
- Z-0661-06
- Recall status
- Terminated
- CDRH record ID
- 44116
- FDA classified
- 2006-03-18
- Recall event number
- 34569