Open original FDA record ↗Recall record
- FDA identifier
- Z-0661-2007
- Company in FDA record
- Sonosite, Inc.
- FDA classification date
- Mar 29, 2007
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
Laparoscopic transducer exceeds the limits specified for radio-frequency emissions by up to 10db.
Source facts
- Action
- On 2/5/2007 firm sent by FED-Ex notification letters to firm representatives. Representatives began hand delivery of recall letter to consignees and retrieval of LAP transducer for return delivery to firm. The letter advised customers that the LAP transducer exceeded the limits specified for radio-frequency emissions. The firm also advised that exceeding the limit may cause the LAP transducer to interfere with surrounding equipment that may be susceptible to radiated emissions and cause surrounding equipment to not perform as expected.
- Initiated
- 2007-02-05
- Root cause
- Other
- Terminated
- 2007-06-01
- Distribution
- Distribution to USA (Tennessee & Ohio), France, Korea, and Germany.
- Recall number
- Z-0661-2007
- Recall status
- Terminated
- CDRH record ID
- 50572
- FDA classified
- 2007-03-29
- Recall event number
- 37452