Open original FDA record ↗Recall record
- FDA identifier
- Z-0661-2008
- Company in FDA record
- Siemens Medical Solutions USA, Inc.
- FDA classification date
- Apr 24, 2008
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
Measurement Error affecting all Sequoia systems using the PAL video standard configuration and have a perspective PC installed: 1) The Sequoia" system is operating at a software revision below 8.0. 2) The Sequoia" system is configured to PAL video mode (normal for systems running at 240V 50 Hz.). 3) While using any of the Perspective" advanced display options listed above, an image or clip is t
Source facts
- Action
- Consignees were notified via a Urgent Medical Device Notification letter/customer letter sent out on 9/25/2007 under Siemens update program US013/07/S. The Siemens Service organization coordinates the implementation of the field correction with the distributors for all countries. Consignees who do not respond to the recall communication will be visit ed and/or telephoned for follow up.
- Initiated
- 2007-09-25
- Root cause
- Software design
- Terminated
- 2010-12-20
- Distribution
- Worldwide Distribution including the countries of: Australia, Austria, Belgium, Brazil, Brunei, China, Denmark, Finland, France, Germany, Great Britain (UK), Greece, Hungary, India, Iran, Ireland, Italy, Latvia, Netherlands, New Zealand, Romania, Russian Federation, Singapore, Spain, Switzerland, Thailand. U.S customers were not affected.
- Recall number
- Z-0661-2008
- Recall status
- Terminated
- CDRH record ID
- 65393
- FDA classified
- 2008-04-24
- Recall event number
- 45410