Skip to content

Public FDA record

Z-0661-2016: Fukuda Denshi patient monitor model DS-8100M and DS-8100N. The Fukuda Denshi DynaScope Model DS-8100N/81

Recall record Z-0661-2016 for Fukuda Denshi USA, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0661-2016
Company in FDA record
Fukuda Denshi USA, Inc.
FDA classification date
Jan 16, 2016
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

A possible malfunction where a Short-term Battery Error message has appeared and has been traced back to a bad (damaged) Lithium-Ion capacitor (Short-term Backup battery).

Source facts

Action
Fukuda Denshi USA, Inc. of Redmond, WA plans to perform 100% field correction of units distributed in the US. The Recall/Field Correction Notice letter (FCN-A01-201512), dated December 21, 2015, and the Customer Confirmation Sheet will be sent to consignees using a clearly red-marked URGENT via Express Mail. The consignees are requested to fill out the Customer Confirmation Sheet and return via fax to 888-224-7090, email to DBlakely@fukuda.com, or send via postal service back to Fukuda Denshi USA, Inc. in a pre-stamped envelope. Consignee can call Doug Blakey directly at 210-240-4521 with any questions or concerns.
Initiated
2015-12-21
Root cause
Process control
Terminated
2016-08-19
Distribution
Distributed in the states of CO, ID, KS, MI, MO, OR, PA, TX, UT, and WI.
Recall number
Z-0661-2016
Recall status
Terminated
CDRH record ID
142581
FDA classified
2016-01-16
Recall event number
72953

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗