Open original FDA record ↗Recall record
- FDA identifier
- Z-0662-2020
- Company in FDA record
- Paragon 28, Inc.
- FDA classification date
- Dec 10, 2019
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
It has been determined that the wedge implant kit is permanently mislabeled with the wrong size. The wedge implant is a 10mm size and is permanently marked as a 12mm size. Use of the product could result in improper preparation of the surgical site causing delays in surgery.
Source facts
- Action
- On September 19th, 2018 Paragon 28 issued a Recall Letter to consignees via email and physical mail. The letter informed consignees of the product mislabeling issue. They were instructed to immediately take the following actions: 1. Review inventory of the TITAN 3D Wedge implants to determine if they possess affected product. 2. If they possess the affected product, return product to Paragon 28 for disposition 3. Regardless, if they have or do not have affected recall product, complete and return an acknowledgement form to Paragon 28
- Initiated
- 2018-09-28
- Root cause
- Error in labeling
- Terminated
- 2020-08-20
- Distribution
- AR, CA, CO, CT, DC, DE,FL, GA, IA, IL, IN, KS, KY, MA, MD, MI, MT, MS, NC, NE, NY, NJ, NV, OH, OR, PA, SC, SD, TN, TX, VT, WA, WI, WY OUS: Australia & Israel
- Recall number
- Z-0662-2020
- Recall status
- Terminated
- CDRH record ID
- 177808
- FDA classified
- 2019-12-10
- Recall event number
- 84234