Z-0662-2025: Exacta External Drainage and Monitoring System: 27581 KIT 27581 EXACTA W/EDM VCATH 35CM 27636 EDM 27636 E

Recall record Z-0662-2025 for Medtronic Neurosurgery. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0662-2025
Company in FDA record
Medtronic Neurosurgery
FDA classification date
Dec 19, 2024
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class I

Reason stated in the source

Due to complaints received reporting cracks and/or leaks of the stopcocks associated with the external drainage and monitoring system (EDMS).

Source facts

Action
On 11/01/2024, the firm sent an URGENT: MEDICAL DEVICE CORRECTION Letter via US mail to customers informing them that through customer complaints, the firm is aware of potential cracking and leaking in the "stopcock" components on the Becker and Exacta External Drainage and Monitoring Systems. Customer are instructed to: Recommended Mitigations: -Prior to use, inspect all stopcocks and connections to ensure that connections are secure and there are no visible cracks in the stopcocks. If cracks or leaks are identified, do not use return the device to Medtronic. -The system must be pre-filled with sterile isotonic saline solution prior to connecting to the patient and all connections checked to ensure that fittings (connections) are tight and leak-free. -All connections should be finger tightened. Over tightening can cause cracks and leaks to occur. -After cleaning with alcohol, or a disinfectant containing alcohol, allow to air dry completely prior to connecting the system. Patient Management Recommendations: -If system develops cracks/leaks during use, replace using sterile technique and return the damaged system to Medtronic. If leaks occur with the Patient Line Stopcock, consider application of a hemostat or other clamp to the proximal patient line while coordinating a replacement if occlusion of the patient line does not create a risk to the patient. -In the product labeling, all patients with EDMS should be monitored for evidence of infection. If cracks or leaks, the system should be replaced, and the patient should continue to be monitored for evidence of infection. If customer chooses to return unused product contact Medtronic Customer Service via email: rs.jaxcustomerservice@medtronic.com, or Phone: 1-800-874-5797 Option 1, and mention FA1452, to initiate return process. For questions-contact Medtronic Rep/Medtronic Technical Support (US) at 1-888-826-5603and reference FA1452.
Initiated
2024-11-01
Root cause
Under Investigation by firm
Distribution
Worldwide distribution. US nationwide including Puerto Rico; Algeria, Argentina, Australia, Austria, Bahrain, Barbados, Bosnia And Herzegovina, Brunei Darussalam, Cambodia, Canada, Chile, China, Colombia, Costa Rica, Curacao, Cyprus, Czech Republic, Denmark, Dominican Republic, Egypt, Finland, French Guiana, French Polynesia, Germany, Ghana, Greece, Guadeloupe, Honduras, Hong Kong, Iceland, Indonesia, Israel, Italy, Jamaica, Jordan, Kazakhstan, Kenya, Kuwait, Lao People s Democratic Republic, Latvia, Lebanon, Libya, Malaysia, Malta, Mauritania, Morocco, Myanmar, Namibia, Netherlands, New Zealand, Nicaragua, Norway, Oman, Panama, Peru, Poland, Portugal, Qatar, Romania, Russian Federation, Rwanda, Saudi Arabia, Singapore, Slovakia, South Africa, Spain, Sweden, Switzerland, Taiwan, United Republic of Tanzania, Thailand, Trinidad And Tobago, Tunisia, Turkmenistan, United Arab Emirates, United Kingdom, Uruguay, Vietnam, and Yemen.
Recall number
Z-0662-2025
Recall status
Open, Classified
CDRH record ID
211018
FDA classified
2024-12-19
Recall event number
95725

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗