Open original FDA record ↗Recall record
- FDA identifier
- Z-0663-2020
- Company in FDA record
- LivaNova USA Inc
- FDA record creation date
- Dec 10, 2019
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
This recall is being initiated due to reports that that the therapy programming tablet with software version 1.5 errantly performs a normal mode diagnostic test instead of the selected system diagnostic test on Model 102 and Model 102R devices, if the output current is greater than 0.5mA. This can result in false high impedance values during patient follow-up.
Source facts
- Action
- On November 16, 2018 LivaNova mailed a "Medical Device Correction" letter via 1st class mail to affected Doctors informing them about the recall. The letter asked doctor's to do the following: 1.When performing System Diagnostics on patients implanted with a Model 102 or 102R generator, do one of the following: a. Use any of the following Programming System software versions: Model 250 8.1/11.0; Model 3000 version 1.0; or b. If one of the above Programming Systems software versions is not available, complete the following steps to perform System Diagnostics using Model 3000 version 1.5 software: i. Interrogate the device, noting the devices current settings, ii. Program normal mode output current to 0mA, iii. Perform System Diagnostics as usual, iv. Re-interrogate the device, v. Program the devices normal mode output current back to the settings noted in step (i) above, and; vi. Interrogate the device to verify settings are as intended prior to the patient leaving the office visit. 2. Please complete and return the attached Customer Response Form by fax to 281-853-1248 or by e-mail to M3000FieldAction@livanova.com. 3. Please ensure that this notice is communicated to all personnel within your organization who need to be aware of it. 4. For questions regarding the information in this letter, please contact Clinical Technical Support at (866) 882-8804 (Monday to Friday, 8 AM to 5 PM CST) or e-mail at cservices@livanova.com.
- Initiated
- 2018-11-16
- Root cause
- Software design
- Distribution
- IL, AK, AL, AR, AZ, CA, CA , CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD MI, MN, MO, MS, MT, NC, ND, NE, NE, NJ, NM, NV, NY, OH, OK, OR, PA, SC, SD, TN, TX, UT, VA, WA, WA , WI, WV, WY,
- Recall number
- Z-0663-2020
- Recall status
- Open, Classified
- CDRH record ID
- 177461
- FDA record created
- 2019-12-10
- Recall event number
- 84170